10
Important Note:
If the deviation is too high, the analyzer is not
adjusted. In this case, please inform your local
Sartorius office, dealer or service center.
§ Open the sample chamber by pressing the
k key
> Allow the moisture analyzer to cool down.
§ Snap the disk holder into the adjustment disk and
remove the disk. Caution: Equipment is extremely
hot after drying!
To avoid severe burns, do not touch until
sufficiently cooled.
!Allow the adjustment disk to cool on a clean,
fire-proof pad.
--------------------
27.05.2009 10:23
Mod. MA100C000230V
Ser. no. 90805350
Ver. no. 01-38-02
ID
--------------------
2-PT.ADJ.:
--------------------
Ref ID 90801234
L.temp. 80'C: 78'C
U.temp. 140'C: 163'C
Adjustment
completed
--------------------
27.05.2009 10:23
Name:
--------------------