7
Adjusting the Temperature
2-Point Adjustment (MA50 and MA100)
Select the 2-point adjustment for samples that
are usually measured at different temperatures.
This adjustment option should also be used when
components such as the temperature sensor or the
sample chamber have to be serviced or replaced.
§ Open the sample chamber by pressing the k key.
§ Remove the disposable pan and pan retainer and
place them on a clean pad.
> ERR 54 or “L” appears on the display
(star-shaped retainer missing).