Short Manual “Calibration” 3.7.01
Calibration
Version 3.7.0 / 2019-06-26 Page 7 of 8
Perform “curvature calibration”:
Move the SIRIUS in front of the calibration
sphere R.8 with radius 8.00 mm.
Align top view image:
The cornea reflex must be
tangential to the red line.
(Image in the superior part
of the window).
Align Placido image:
The red cross must be in the
center of the placido rings
(image in the inferior part of
the window).
When both images are in the correct position, press the joystick button.
When the calibration was successful the
following window appears:
Click “OK” to confirm the calibration.
If the calibration was unsuccessful, repeat the
whole calibration procedure.