Short Manual “Calibration” 3.7.01
Calibration
Version 3.7.0 / 2019-06-26 Page 8 of 8
At the end measure the curvature sphere of
the calibration tool once and compare the
achieved average value of the K-readings
Meridians @3mm with the value of the
calibration sphere. The average (Avg) shall be
8.00mm±0.03mm (42.2D±0.16D).
If the measured average K-value is out of
range repeat the calibration procedure.
Finally check acquisition quality, the keratoscopy centration shall be 100% ± 3%.
Now the instrument is ready for use.