SciAps X-550 and X-505 QuickStart Guide
6
3. Calibrating the unit
If this is the first time the unit is being used (or the first time the
unit being used for the day), the unit will need calibrating.
If calibration is required, the unit will automatically bring up the
Calibration window.
Place the 316 coupon that came with the unit on the nose.
If the coupon is not available, use a clean, known 316 steel
sample.
Make sure the 316 coupon completely covers the analysis
window
Tap Calibrate to begin the process.
Calibration takes 15 seconds.
1 of 15 means the first second of the 15 seconds of the
calibration.