EasyManua.ls Logo

Scion Instruments 456-GC - Instrument Information and Compliance; Understanding Safety Indications

Scion Instruments 456-GC
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Scion Instruments
Scion Instruments Natgas C Analyzer +He/H
2
Channel Reference Manual Page: 7
OPERATION
An operational method is preloaded in the instrument.
Listed below is an example method; settings determining the application
may be different.
Sample delivery
V1 (GSV)
V2 (CS)
V4 (S)
Initial
-
-
-
0.01
- - +
0.80
-
-
-
1.00
+
-
-
1.91
+ + -
Injector (S/SL Front):
Temp 220 ºC
Split ratio 200
Flow/Pressure:
Front EFC-25:
Pressure (PSI)
Hold (min)
20.0
18.00
Middle EFC-24:
Pressure (PSI)
Hold (min)
14.0
18.00
Total Flow: 60 ml/min
Rear EFC-24:
Pressure (PSI)
Hold (min)
12.0
18.00
Total Flow: 60 ml/min
Column Oven:
Temp(ºC)
Rate(ºC/min)
Hold(min)
Total(min)
Initial
45
--
6.00
6.00
190
20
4.75
18.00
Oven stabilization time 0.50 minutes
The temperature is raised after elution of H
2
S.
Large Valve Oven:
Temp 100 ºC

Other manuals for Scion Instruments 456-GC

Related product manuals