4.4 Clearing stored measurement data
In memory mode, you can delete all the measurement data of the two users by pressing the
“MEM” button for 3 seconds.
4.5 Use Bluetooth function
See the User manual of Jumper Health.
4.6 Classification standard for blood pressure condition
There is no definition of hypotension yet. Generally, if the systolic blood pressure is less
than 90mmHg (12kPa), it is hypotension
CONTRAINDICATIONS, PRECAUTIONS, WARNING AND
INSTRUCTION
(1) Warning:
Keep the device out of the reach of the children under 12 and people who can’t express
their intention. When children of 12~18 use the device, they should be accompanied by the
adults. Pregnant women shall use under the guidance of doctors.
Do not wrap the cuff over a wound, as this can cause further injury.
This model of Electronic Blood Pressure Monitor is suitable for the arm circumference range
of 22-36cm or 22-42cm, and if the arm circumference exceeds this range, you might fail to
obtain the correct measured value of blood pressure;
This model of Electronic Blood Pressure Monitor is not suitable for newborns or young
children;
The blood pressure is constantly changing. You shall not judge the blood pressure condition
with just one measurement result. The repeated measurement data over a period of time
will be more reliable;
Do not make self-diagnosis according to the measurement results. Please consult your
professional doctor with the measurement result record(s). The treatment based on the self-
diagnosis of measurement results is very dangerous;
Do not collide or drop the host to avoid collision or strong impact on the device;
Do not mix old and new batteries of different brands for use.
Indoor temperature, environment, noise, user's body position, speech or exercise might
affect the blood pressure measurements. The body movement, magnetic field and improper
wearing of cuff and sensor will affect the measurement as well.
TROUBLESHOOTING
* If you have trouble using the device, please check the following instruction.
STORAGE AND MAINTENANCE
Keep the device away from direct sunlight, extreme temperatures, humidity or
moisture.
Use a dry, soft cloth to clean the device, or if desired, use a cloth lightly
dampened with water.
Do not use corrosive cleaner, benzene, thinner or other volatile liquids to
clean the device.
Do not wash or expose the arm cuff to liquid.
Remove batteries from the device when it will not be used for more than 3
months.
Disinfection:
Recommended disinfecting agent:
Isopropanol solution with 70% concentration
Medical alcohol with 75% concentration
SPECIFICATIONS
Problem
Possible cause
How to correct
No display when
press the Power
button
Low battery
Replace new batteries or use
the USB for power supply.
The polarities of
batteries are installed
wrongly.
Install the batteries in correct
polarities.
No pressurizing
The air plug is
loosely installed.
Make sure the air plug is
securely inserted in the main
unit.
The air tube is
broken or leaked.
Purchase a new cuff.
“Err 1” displayed
The sensor doesn’t
work.
Please contact us for after-sale
service.
“Err 2” displayed
No pulse detected
Please check if the cuff wearing
is correct, and follow Section 4.1
How to use the cuff
“ Err 3
displayed
The result of blood
pressure is
abnormal.
Please remeasure in the right
way.
“Err 4” displayed
The cuff is twined too
loose.
Please check if the cuff wearing
is correct, and follow Section 4.1
How to use the cuff.
“Err 5” displayed
The air bag of cuff is
ruptured.
Replace with a new cuff.
“Err 6” displayed
The air tube is
detected to be
obstructed in the
early stage of
pressurization.
Please check if the air tube is
clogged or folded before
measuring
Please measure again.
“Err 7” displayed
The air bag of cuff is
ruptured.
Replace with a new cuff.
“Err 8” displayed
Pressure value
exceeds the limit.
Don’t touch the cuff during the
measurement;
Don’t fold the air tube during the
measurement.
Note: If your problem cannot be solved by the above, please
contact customer service. Do not disassemble the device!
Product Name
Electronic Blood Pressure Monitor
Model
JPD-HA120
Display Mode
Digital display mode
Measuring Mode
Oscillographic method
Measuring Body Part
Upper arm
Measuring Range
Pressure value
0-295mmHg0kPa-
39.3kPa
Pulse value
40-199 pulse beats/min
Static measurement
accuracy
Pressure value
±3mmHg±0.4kPa
Pulse value
± 5% of read value
LCD display
Pressure
Unit : mmHg/ kPa
Pulse
Pulse rate per minute,
displaying three digits
Storage capacity
It can store 99 groups of measurement data for each of two
users
Power supply
4 AA dry batteries
/DC 5V USB external power supply
Power Off Mode
Manual power-off/ Auto power-off
Device weight
(without batteries)
About 285g