EMC Information-Guidance and Manufacture’s Declaration
1* WARNING: Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating
normally.”
2* WARNING: Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in improper
operation.”
3* WARNING: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the ME equipment, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.”
Table 1
declaration - electromagnetic emission
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
declaration - electromagnetic immunity
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 0.5kV, ± 1 kV line(s) to lines
± 0.5kV, ± 1 kV, ± 2 kV line(s) to
earth
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0 % UT; 0.5 cycle At 0°, 45°,
90°, 135°, 180°, 225°, 270°and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level
declaration - electromagnetic immunity
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM bands between 0.15
MHz and 80 MHz
Radiated RF
IEC 61000-4-3
declaration - IMMUNITY to proximity fields from RF wireless communications
equipment
Radi
ated
RF
IEC
6100
0-4-3
*FM+ 5Hz
deviation:
1kHz sine
**Pulse
Modulation:
217Hz
1720 MHz
1845 MHz
1970 MHz
**Pulse
Modulation:
217Hz
**Pulse
Modulation:
217Hz
5240 MHz
5500 MHz
5785 MHz
**Pulse
Modulation:
217Hz
Note* - As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be worst case.
Note** - The carrier shall be modulated using a 50 % duty cycle square wave signal.
FCC Compliance Statements
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) this device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.
Note: This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a
particular installation.
If this equipment does cause harmful interference to radio or television reception, which
can be determined by turning the equipment off and on, the user is encouraged to try to
correct the interference by one or more of the following measures:
-- Reorient or relocate the receiving antenna.
-- Increase the separation between the equipment and receiver.
-- Connect the equipment into an outlet on a circuit different from that to which the receiver
is connected.
-- Consult the dealer or an experienced radio/TV technician for help.
Caution: Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
US Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled
environment. End user must follow the specific operating instructions for satisfying RF
exposure compliance. This transmitter must not be co-located or operating in conjunction
with any other antenna or transmitter.
After-sale Service
After-sale service unit: Shenzhen Jumper Medical Equipment Co., Ltd. Address: D
Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, China Tel:
+86-755-26696279 E-mail: info@jumper-medical.com
Website: www.jumper-medical.com
MedPath GmbH
Mies-van-der-Rohe-Strasse 8, 80807 Munich, Germany
Shenzhen Jumper Medical Equipment Co., Ltd.
D Building, No. 71, Xintian Road, Fuyong Street,
Baoan, Shenzhen, Guangdong, China,518103
Tel: +86-755-26696279
E-mail: info@jumper-medical.com
Website:www.jumper-medical.com
Distributed by:
Screen Italia Srl
Via dell’Artigianato, 16
06089 - Torgiano - Perugia - Italia
www.screenpharma.it - info@screenpharma.it
138mm (length) *120mm (width) * 59mm (height)
60mm (length) *77mm (width)
Upper-arm-type cuff
(measured arm circumference 22-36cm or 22-42cm)
Cuff, User Manual, Dry battery, Zipped Carrying Bag
High-performance dry battery can be used for about 300
times at normal temperature
If the device is
stored and
used in the
environment
out of the
designated
temperature &
humidity
ranges, it
cannot operate
normally.
Temperature Condition 5 ℃-40 ℃
Humidity Condition 15%-85%RH
Atmospheric Condition 70kPa-106kPa
Transportation and
Storage
Environment
Avoid strong impact, direct impact, exposure or rain during
transportation. The packaged Blood Pressure Monitor shall
be stored indoors at the temperature of -20℃~55℃ and the
relative humidity of 10%~93%, atmospheric
Condition70kPa-106kPa, without corrosive gas and with
good ventilation