EasyManua.ls Logo

Seca 769 LC - Description of Faults; Troubleshooting Scale Malfunctions

Seca 769 LC
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual Description of faults
19.07.07 /Kr/Rei 30-34-00-744
Scales with LC electronic
Please refer to the operating instructions to ensure that the scales were operated in the
proper manner. In case of a malfunction, the following table will help you to locate the fault.
Description of the fault
Possible cause Remedy
Scales cannot be started • Battery is not
installed or
discharged
• Check the batteries
• Voltage supply is
defective
• Check the voltage at the appropriate connections of the electronics;
eliminate the interruption
• Bus connection
cable is defective
• Check the plug-in connections; check the cables for interruption
and replace them if necessary.
• Keypad is defective
or not connected
• Check the cable connections; check the cables and the keypad for
interruption and replace them if necessary
• Bus fault • Disconnect power supply unit and battery from the scale, wait for 60
seconds, then reconnect.
Scales show no weight or
an incorrect weight
(repeatedly)
• Packaging /
transport safeguard
has not been fully
removed
• Check the scales
• Scales are
maladjusted
• Recalibrate the scales
• Sensor is damaged • Replace the sensor and start an additional corner override
transfer of force to the
sensor is interrupted
• Check the base frame for damage or incorrect assembly
Scale shows only the
segment test or the
readings are varying to
much.
• Shortcircuit between
sensor cable and
housing
• Search for damaged isolation of the cabel and repair it. ( If
appropriated use multimeter.)
Incomplete digits • Display is defective • Replace the electronics
Malfunction during the
weighing procedure
• Display:
Er:x:no
• The electronics have detected an error.
n X: module number
n No: error number
E XX.11
•Sensor is defective or
supply lines are
damaged
• Check connection cables, check sensor and replace as required (if
parts a replaced, a corner override is necessary)
• If several load cell are used, check the allocation (colour codes)
E XX.12
•Failure at
Identification
initial zero setting
• Check the initial load
• Sensor defective, replace it possibly (a corner override is
necessary)
E XX.13
• Limit for zero
tracking is reached
• Load the scale at work with >50g/s
E XX.16
•Failure at
Identification
initial zero setting
• while the start process beware the scale for agitation
E XX.40
• No EEProm data • Recalibrate the scale
E XX.50
• Flash memory
damaged
• Replace the electronics
E XX.60
• Bus connection
cable is defective
• Check the plug-in connections; check the cables for interruption and
replace them if necessary, Disconnect power supply unit and
battery from the scale, wait for 60 seconds, then reconnect.