Christopher HowardAug 15, 2025
What to do if Sejoy BG-203 battery power is low?
- DDavid FrenchAug 15, 2025
If the Sejoy Blood Glucose Meter displays that the battery power is low, replace the battery soon.
What to do if Sejoy BG-203 battery power is low?
If the Sejoy Blood Glucose Meter displays that the battery power is low, replace the battery soon.
What to do if the sample is applied after the Sejoy Blood Glucose Meter times out?
If the sample was applied after the Sejoy Blood Glucose Meter timed out and turned off, remove the test strip and repeat the test using a new one. Ensure you wait until you see the blood and test strip symbols on the display screen before applying the blood sample.
What to do if the test strip moves during testing with Sejoy BG-203 Blood Glucose Meter?
If the test strip may be moved during testing or sampling data is unstable, retest with a new strip. Ensure the method of applying the sample is correct, and the test strip remains still during testing.
What to do if the sample was applied before the Sejoy Blood Glucose Meter was ready?
If the sample was applied before the Sejoy Blood Glucose Meter was ready, repeat the test with a new strip. Make sure to apply blood after the symbol flashes on the screen.
What to do if Sejoy BG-203 Blood Glucose Meter is out of operating temperature range?
If the Sejoy Blood Glucose Meter is out of operating temperature range, place the system in an appropriate operating environment for 30 minutes before retesting.
What to do if test strip inserted upside down in Sejoy Blood Glucose Meter?
If the test strip is inserted upside down, or incompletely inserted into the Sejoy Blood Glucose Meter, check that the test strip is inserted correctly.
How to fix Sejoy Blood Glucose Meter system check failed?
If the Sejoy Blood Glucose Meter system check failed, remove the battery and reinsert it after 30 seconds. If the problem persists, it is recommended to contact the vendor.
What to do if Sejoy BG-203 Blood Glucose Meter strip check has problem?
If the Sejoy Blood Glucose Meter shows a strip check problem, make sure that the strip model is correct and retest with a new strip.
What to do if the test strip is damaged for Sejoy Blood Glucose Meter?
If the test strip may be used or damaged, make sure that the strip model is correct and retest with a new strip.
What to do if Defective test strip for Sejoy BG-203?
If you suspect a defective test strip with your Sejoy Blood Glucose Meter, repeat the test with a new test strip.
Provides crucial safety precautions and warnings for using the blood glucose meter and lancet device.
Advisories on health conditions affecting test results and treatment decisions.
Step-by-step guide to configure the meter's date and time settings.
Detailed instructions on how to prepare the lancing device for blood sampling.
Procedures for obtaining a blood sample from the fingertip.
Steps involved in inserting a test strip and applying the blood sample for measurement.
Guidance on understanding 'Lo', 'HI' readings and other unusual test outcomes.
Step-by-step guide on how to use control solution with the meter.
Identifies common error codes displayed by the meter and their corresponding solutions.
| Blood Sample | Capillary Whole Blood |
|---|---|
| Sample Type | Fresh capillary whole blood |
| Measuring Time | 5 seconds |
| Measuring Range | 20-600 mg/dL |
| Sample Volume | 0.5 µL |
| Memory | 300 test results |
| Operating Temperature | 10-40°C |
| Battery | One 3.0V CR2032 button cell battery |
| Automatic Shut-off | 2 minutes after last action |