INSTRUCTION MANUAL CODE 80152 REV E DEC 2019 (They can be modied without notice) Page: 17
J.P. SELECTA s.a.u. Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: www.grupo-selecta.com
Estados de alarma
ALARMA CONCEPTO RESPUESTA
0 Ninguna alarma #NNALARM0KK<CR><LF>
1 Sobretemperatura #NNALARM1KK<CR><LF>
2 Fallo de red #NNALARM3KK<CR><LF>
Comandos de lectura
COMANDO CONCEPTO MENSAJE RESPUESTA
PVT? Temperatura PV (ºC) #NNPVT?KK<CR><LF> #NN+000KK<CR><LF>
PVP? Presión PV (mbar) #NNPVP?KK<CR><LF> #NN0000KK<CR><LF>
SVT? Set temperatura(ºC) #NNPVT?KK<CR><LF> #NN+000KK<CR><LF>
SVP? Set presión (mbar) #NNSVP?KK<CR><LF> #NN0000KK<CR><LF>
RUN? Estado aparato #NNRUN?KK<CR><LF> #NNRUNKK<CR><LF>
ID? Identidad aparato #NNID?KK<CR><LF> #NN>KKKKKKKVKK<CR><LF><CR><LF>
SAL? Estado alarmas #NNSAL?KK<CR><LF> #NNALARM0KK<CR><LF>
TBY? Tiempo standby #NNTBY?KK<CR><LF> #NN00,0KK<CR><LF>
TBR? Tiempo espera marcha #NNTBR?KK<CR><LF> #NN00,0KK<CR><LF>
TRU? Tiempo de marcha #NNTRU?KK<CR><LF> #NN0,00KK<CR><LF>