62 | All-Day Clear Slim All-Day Clear Slim | 63
Medical examinations with MRI/NMRI scans,
generating magnetic fields.
Hearing aids don’t need to be removed when
passing security gates (airports etc.). If X-ray
is used at all, it will be in very low doses, and
will not affect the hearing aids.
Patients implanted with magnetic externally
programmable CSF shunt valves may have a
risk of experiencing an unintended change in
their valve setting when exposed to strong
magnetic fields. The receiver (loudspeaker)
of hearing aids contain static magnets. Keep
a distance of 5 cm (approx. 2 inches) or more
between the magnets and the site of the
implanted shunt valve.
Use of this equipment adjacent to or stacked
with other equipment should be avoided
because it could result in improper operation.
If such use is necessary, this equipment and
the other equipment should be observed to
verify that they are operating normally.
Only use the charging case and power supply
provided by the manufacturer for charging.
Otherwise, the device(s) may be damaged.
Use of accessories, transducers and cables
other than those specified or provided by the
manufacturer of this equipment could result
in increased electromagnetic emissions or
decreased electromagnetic immunity of this
equipment and result in improper operation.
Your hearing aids operate in the frequency
range of 2.4 GHz–2.48 GHz. When flying
please check if flight operator requires
devices to be switched into flight mode
(more information can be found on the
product support webpage).
Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches) to any
part of the hearing aids, including cables
specified by the manufacturer. Otherwise, the
equipment’s performance could be degraded.
The device has not been tested for known
sources of electromagnetic interference
such as wireless power transfer systems,