EasyManua.ls Logo

Sensus FM-1D/K - Protection; Device Protection Mode; Test Mode; Activating Test Mode

Default Icon
5 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
the lower status LED blinks the last 4 digits can be set.
The setting is identical as described in point 1. The posi-
tion of the decimal point depends on the output pulse
value ( see point 4)
3. Pulser
The setting of the pulse value of the sensor in litre per
pulse. In this example with 1 litre per pulse the display is
to be set to ”0001.”. The possible input pulse range is
0.001 ... 9999. litres/pulse. After the pulse value pro-
gramm the signal mode suitable for the connected pulser.
InPu= Opto OD 01-04 without evaluation of the direction
signal
InPb= Opto OD 01-04 with evaluation of the direction sig-
nal
InPr= Opto OD 01-04 without direction signal
4. Pulse output
For example to set 1 pulse per m
3
, the display must indi-
cate ”0001.”. The setting is identical as described in
point 1. The possible output pulse value is 0.001 ...
9999.m
3
. Note ! The output pulse value has to be > to the
input pulse value. At max. flowrate the output pulse rate
must not exceed 1 Hz.
5. Current output
The actual current output range will be indicated at the
display. With the (+) or (-)-button it can be toggled
between 0...20mA and 4...20mA. Pressing the SELECT
button saves the setting.
6. Damping
The damping determines the response time of the cur-
rent output and the flowrate indication.
If there are no special requirements for the response
time please set damping value ”0004”.
Setting 1 - no damping Setting 14 - max. damping
*) from serial no. 105.000
5. Protection
The FM-1D/K can be protected against unintentional
changes to the setting. In the protect-mode all settings
and measuring values can be displayed. Changing the
values is not possible.
Simultaneously by pressing (+) and (-)-buttons and the
SELECT-button, the present mode will be displayed.
display ”LOC 6”
- device in programming mode,
display ”LOC 8”
- device protected.
With the (+) or (-)-buttons modes can be toggled
between ”LOC 6” and ”LOC 8”. Pressing the SELECT
button saves the setting.
6. Test mode
In the test-mode the digital and ana-
logue output of the FM-1D/K can be
checked on the integral counter and flowrate indicator,
without having a pulser connected to the pulse input.
First the firmware revision will be indicated.
By pressing the (+) or (-)-buttons the test mode starts:
The current output toggles, as indicated on the display,
in a 30 second interval between 0(4) und 20 mA. The
flowrate indicator should alternate between ”0” and full
scale.
At the pulse output terminals a pulse will be generated
every 60 seconds. These pulses will be indicated on
the pulse counter.
Pressing the SELECT-button leaves the test mode and
changes into the standard mode.
Test
I in mA
t in sec.
Pulse
Pulse
30 60 90
20
4
120