Sentec AG, Ringstrasse 39, CH-4106 Therwil, Switzerland, www.sentec.com
HB-010103-e Date of release: 01/2022
2
Required & recommended components for setup
REF: RFT100VA-V1 / RFT100VA-V2
Sentec Isolation Transformer
(see chapter 3.2 for technical data)
IP42
Water Protection Box (not included)
IP44
Note: Recommended - a similar product with a minimum
degree of IPX2 (ingress protection) may be used.
SDM Mains Cable (enclosed with SDM)
Caution: Use only hospital-grade power cords provided by
Sentec.
Note: Mains cables are country specific, and may differ in
shape and/or color.
This manual provides systematic guidance for an instructed person from a homecare provider on setting up the
SDMS at the patient’s home/site. The SDMS shall only be used if all of the following steps have been carried out
successfully.
Note: Installation of the SDMS including configuration of the monitor (SDM) shall only be
performed by an instructed person from a homecare provider. This instructed person must
have received the appropriate training by a Sentec AG representative. SDMS-related tutorials
are available for online viewing at http://www.sentec.com/tv .