8
Country specific example (CEE 7/7)
Plug the SDM mains cable into the wall socket. Do
not use extension cables or multipliers.
Ensure that the mains connection can be easily
disconnected from the mains supply.
Note: As an additional means of protection,
Sentec recommends to perform a socket and
wiring test prior to use (see Annex).
Circuit
In case the circuit breakers on the back of the
Isolation Transformer are triggered, they can be
reset by pushing them back into their lock position.
Note: The Isolation Transformer has an overheat
protection inside, which responds at approx. 120°C
and switches off the transformer. The overheat
protection is self-resetting and will only switch on
the transformer again when it has cooled down.
Simplified depiction of final SDM & Isolation Transformer assembly
Instructed personnel from the homecare provider shall electrically test the Isolation Transformer at
least every 6 months (chapter 3.1).
Refer to the Instruction Manual for the SDMS (HB-005771) or Technical Manual for the SDMS (HB-005752)
for additional information, e.g. transport and storage conditions.
Troubleshooting is provided in the ‘Service Manual for the SDMS’ (HB-005615)
available on www.sentec.com/ifu.
Leave this Instruction Manual (HB-010103) at the patient’s home during monitoring
(also available on www.sentec.com/ifu
).
Provide the Lay User Manual (HB-010069-Directions for Lay Users) to the patient (also available on