SPS-1500 Troubleshooting
12
Error
Possible Cause
Troubleshooting Step
No Tool
(No Tool indicates there is
no gauge current)
Blown SPS Fuse
Blown Surge Protector Fuse/MOV
Bad SPS to Gauge Signal wire
connection
No-Tool Current setting incorrect
in SPS
Read the DC current drawn by the gauge with an ammeter in
series and DC mode. DC current is normally 15-30mA.
Check all fuses and surface connections.
If connections and equipment okay error may indicate possible
downhole connection is open.
Using “DataWorks”, verify “No Tool Amp” value is set to 5.3mA
(default value).
No Tool
Indica que no hay sensor
conectado
Fusible SPS abierto
Fusible / MOV del protector de
sobretensión abierto
Mala conexión del cable de SPS a
la señal del sensor
Configuración de corriente No-
Tool Current incorrecta en SPS
Lea la corriente DC consumida por el sensor con un
amperímetro en serie y modo DC. La corriente DC debe ser
normalmente de 15 a 30 mA.
Compruebe todos los fusibles y las conexiones de superficie.
Si las conexiones y el equipo están bien, el error puede indicar
que una posible desconexión en el fondo.
Usando “DataWorks”, verifique que el valor de “No Tool Amp
esté establecido en 5.3mA (valor predeterminado).