DO NOT USE CHAIR UNTIL YOU HAVE FULLY READ THE INSTRUCTIONS
5.5 LINAK OPENBUS
DO’S and DON’TS
5.5.1 Before connecting or disconnecting OPENBUS™ accessories or handsets to / from the system:
a.
Disconnect the CB from mains power.
b.
Wait at least 2 minutes after the last key press.
5.5.2
Only connect the originally specied type of Actuator to the appropriate port on the CB.
5.5.3
Only connect the originally specied Handset to the system.
5.5.4
Wait at least 2 minutes after the last key has been pressed before removing the BA18 or BA20 battery from
your system. Otherwise battery operation may be blocked for 30 minutes.
5.5.5
Do not connect / disconnect an Actuator while pressing any keys on the Handset or Foot Switch.
5.5.6
Use O-rings on all connectors (greasing them 1st) and secure all connectors / cables with the cable locking
devices supplied.
5.5.7
Use blind plugs with O-rings to seal any unused ports on the CB and junction boxes.
5.5.8
Do not stretch or unduly strain any of the cables and connectors.
5.5.9
In cases where a CU20 or CP20 requires replacement, in order to avoid damage to new components, replace
the complete CB20, rather than individual components.
5.5.10
Ensure that the system is suitably protected from ESD (Electro Static Discharge) via the ESD drag chain/wire
tted to the base of your chair. If worn out, replace with an original part.
5.5.11
In the event of a breakdown of the chair electric functions, do not diagnose the fault by switching actuators
around between dierent working ports, or use components from a working chair on the faulty chair. This
system is not a ‘Plug and Play’ type. It is computerised and sensitive to faulty parts. A faulty or damaged
component may cause ruin/damage to a working/new component. It is always best to call a technician with a
Linak approved repairer qualication to diagnose the issue, if it is not obvious.
5.5.12
In the event components are replaced, it is always best to perform a Reset and System Initialisation after doing
so, to enable correct function of the chair – refer to 5.7.1 and 5.7.2 on Page 21