2223A Oxygen Transmitter Functional Safety Manual
3.4 Settings
After installation and commissioning according to the installation manual, the following
settings should be made to support the safety function. This will require the use of a 2210
or 2213 Control Unit or a 2215 Tx Interface. To access these functions press the Enter key,
navigate to
Passwords
→
Enter Password
and enter the appropriate password to allow settings to be changed.
From the initial menu screen, navigate as follows:-
Transmitters
→
Transmitter Setup
→
Select Transmitter
→
Peripherals
Enter the appropriate forms and ensure that the necessary settings are made as described
below. Note that it may be necessary to scroll the forms displayed on the control unit in
order to see all of the fields shown in the following sections.
3.4.1 mA Output
Configure the mA output as shown as described below:
• Measurement 1 and 2 should be set to appropriate values for the application.
• Under Range Current must be a non-zero value.
• Jam must be set to Low or High. If the equipment to which the mA output is
connected does not recognise
both 0mA and 21mA as an indication of fault, then
this parameter must be set to Low.
• On Service In Progress should be set as required.
8 02223006A / Revision 0