Specifies that Model 264 transducers are for air or nonconducting gases only, warning against liquids or corrosive gases.
Details the operating and compensated temperature limits for the Model 264 transducer.
Describes recommended tubing sizes (3/16", 1/4", 3/8" I.D.) for different lengths up to 900 feet.
Explains the 3-wire circuit for voltage output units (COM, EXC, OUT) and excitation requirements.
Details the 2-wire loop-powered 4-20mA output, terminal designations, and power supply requirements.
Describes how to adjust the zero output for voltage units while monitoring the output signal.
Explains how to adjust the span for voltage units using a pressure standard.
Details how to adjust the zero output for current units by monitoring current.
Describes how to adjust the span for current units using a pressure standard.
Specifies that Model 264 transducers are for air or nonconducting gases only, warning against liquids or corrosive gases.
Details the operating and compensated temperature limits for the Model 264 transducer.
Describes recommended tubing sizes (3/16", 1/4", 3/8" I.D.) for different lengths up to 900 feet.
Explains the 3-wire circuit for voltage output units (COM, EXC, OUT) and excitation requirements.
Details the 2-wire loop-powered 4-20mA output, terminal designations, and power supply requirements.
Describes how to adjust the zero output for voltage units while monitoring the output signal.
Explains how to adjust the span for voltage units using a pressure standard.
Details how to adjust the zero output for current units by monitoring current.
Describes how to adjust the span for current units using a pressure standard.
| Output Signal | 4-20 mA, 0-5 VDC, 0-10 VDC |
|---|---|
| Power Supply | 12-30 VDC |
| Accuracy | ±0.25% FS |
| Media Compatibility | Non-corrosive gases |
| Pressure Port | 1/4" NPT |