O P E R A T I O N G U I D E
OPERATION GUIDE
M517-E077A

Questions and Answers

  • K
    kathryn24Sep 23, 2025
    What actions are needed after a power failure with Shimadzu Sonialvision Safire?
    • D
      Danielle HamiltonSep 23, 2025
      Following a power failure with the Shimadzu Diagnostic Equipment, perform a manual startup.
  • M
    Mark BishopAug 23, 2025
    What to do if Shimadzu Sonialvision Safire Diagnostic Equipment displays error 'HU Full Stored'?
    • C
      Christopher MaxwellAug 23, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'HU Full Stored,' it indicates that the heat units have reached the permitted limit. You should cease operation until the heat unit value reduces.
  • D
    Donna Holt MDAug 25, 2025
    What to do if Shimadzu Sonialvision Safire displays error 'Thermal OVER'?
    • J
      Julie ChambersAug 26, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'Thermal OVER,' it means the X-ray tube unit temperature has exceeded the permitted limit. To resolve this, cease operation until the heat unit value reduces.
  • D
    Dawn DiazAug 27, 2025
    What to do if Shimadzu Diagnostic Equipment shows error 'Radio Over Current'?
    • W
      willisryanAug 27, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'Radio Over Current', it signifies that the radiography tube current exceeded the set value + 200 mA. To resolve this, change the set value.
  • T
    travis57Aug 27, 2025
    What to do if Shimadzu Diagnostic Equipment shows error 'Emission OVER'?
    • M
      Michael HillAug 28, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'Emission OVER,' it indicates that the emission characteristics are out of range. You can resolve this by reducing the set value of mAs or sec.
  • J
    jonesgabrielSep 17, 2025
    What to do if Shimadzu Diagnostic Equipment displays error 'Position Error XXX'?
    • B
      brianrussoSep 18, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'Position Error XXX,' it indicates a potentiometer break detect. To resolve this, reset by pushing the STOP switch, or re-startup the system.
  • S
    Samuel HendricksSep 14, 2025
    What to do if Shimadzu Sonialvision Safire Diagnostic Equipment displays error 'The fluoroscopy table is not prepared'?
    • T
      Thomas CraigSep 14, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'The fluoroscopy table is not prepared,' it means the fluoroscopy table is not prepared due to the incorrect position of the X-ray tube. To resolve this, confirm the diagnostic table configuration.
  • A
    Ariel WilliamsSep 15, 2025
    What to do if Shimadzu Sonialvision Safire Diagnostic Equipment displays error 'Position Detect Error XXX'?
    • J
      Jeffrey GrossSep 16, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'Position Detect Error XXX,' it indicates a position detect error in the operating parts. To resolve this, reset by pushing the STOP switch, or re-startup the system.
  • A
    Amanda RileySep 6, 2025
    What to do if Shimadzu Diagnostic Equipment shows error 'I.F. OVER'?
    • B
      Beverly NicholsSep 6, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'I.F. OVER,' it means there is an abnormal filament heating current. To resolve this, reduce the set value.
  • M
    Ms. Hailey NavarroSep 3, 2025
    What to do if Shimadzu Sonialvision Safire Diagnostic Equipment shows error 'Starter ERR'?
    • G
      gary72Sep 3, 2025
      If the Shimadzu Diagnostic Equipment displays the error message 'Starter ERR,' it indicates the starter is in an abnormal situation or condition. To resolve this, decrease the radiography tube voltage or radiography tube current.

Summary

System Overview and Components

System Configuration

Details the overall system components and layout.

Diagnostic Table Console Operation

Explains the controls and functions of the diagnostic table console.

X-ray Generator Console Operation

Describes the operation and controls of the X-ray high voltage generator.

Image Processor Main Screen

Outlines the main display and controls of the digital image processor.

System Operation Workflow

Provides a visual guide to the system's operational sequence.

System Startup and Shutdown Procedures

System Startup Process

Step-by-step guide for powering on the system and its components.

System Shutdown Process

Instructions for properly closing active studies and powering off the system.

FPD Calibration Procedure

Procedures for calibrating the Flat Panel Detector (FPD) to ensure image quality.

FPD Calibration via Right-Click Menu

Alternative method to perform FPD calibration using the system's context menu.

Initiating Patient Studies

Entering Patient Information for Usual Study

Process for entering patient information to initiate a new study.

Emergency Patient Data Entry

Method for quickly entering patient data during emergency situations.

Selecting Study Modalities

How to select the appropriate study type based on imaging needs.

Fluoroscopy and Radiography Operations

Setting Imaging Conditions

Steps to configure parameters for radiography and fluoroscopy procedures.

SPOT and SERIAL Radiography

Detailed procedure for performing SPOT and SERIAL radiography.

DSA Radiography (Optional)

Instructions for performing Digital Subtraction Angiography (DSA) if the option is available.

Digital Tomography (Optional)

Guide for operating the digital tomography function, including parameter setup.

Image Processing and Management

Printing Acquired Images

Methods for printing acquired images from the system.

Printing Images via Right-Click Menu

Procedures for printing specific images using the context menu.

Printing Images from Patient List

How to print entire patient files or selected studies from the patient list.

DICOM Image and Study Transmission

Instructions for transmitting images and studies using DICOM network or CD-R.

Closing Patient Studies

Closing the Active Study Session

Steps to properly close the current patient study.

Troubleshooting and Error Handling

Emergency Stop and System Recovery

Procedures for using emergency stop and recovering the system.

System Error Messages and Solutions

Lists common error messages from the X-ray generator and their solutions.

Post Power Failure Recovery Actions

Steps to take for system recovery and verification after a power interruption.