What actions are needed after a power failure with Shimadzu Sonialvision Safire?
- DDanielle HamiltonSep 23, 2025
Following a power failure with the Shimadzu Diagnostic Equipment, perform a manual startup.
What actions are needed after a power failure with Shimadzu Sonialvision Safire?
Following a power failure with the Shimadzu Diagnostic Equipment, perform a manual startup.
What to do if Shimadzu Sonialvision Safire Diagnostic Equipment displays error 'HU Full Stored'?
If the Shimadzu Diagnostic Equipment displays the error message 'HU Full Stored,' it indicates that the heat units have reached the permitted limit. You should cease operation until the heat unit value reduces.
What to do if Shimadzu Sonialvision Safire displays error 'Thermal OVER'?
If the Shimadzu Diagnostic Equipment displays the error message 'Thermal OVER,' it means the X-ray tube unit temperature has exceeded the permitted limit. To resolve this, cease operation until the heat unit value reduces.
What to do if Shimadzu Diagnostic Equipment shows error 'Radio Over Current'?
If the Shimadzu Diagnostic Equipment displays the error message 'Radio Over Current', it signifies that the radiography tube current exceeded the set value + 200 mA. To resolve this, change the set value.
What to do if Shimadzu Diagnostic Equipment shows error 'Emission OVER'?
If the Shimadzu Diagnostic Equipment displays the error message 'Emission OVER,' it indicates that the emission characteristics are out of range. You can resolve this by reducing the set value of mAs or sec.
What to do if Shimadzu Diagnostic Equipment displays error 'Position Error XXX'?
If the Shimadzu Diagnostic Equipment displays the error message 'Position Error XXX,' it indicates a potentiometer break detect. To resolve this, reset by pushing the STOP switch, or re-startup the system.
What to do if Shimadzu Sonialvision Safire Diagnostic Equipment displays error 'The fluoroscopy table is not prepared'?
If the Shimadzu Diagnostic Equipment displays the error message 'The fluoroscopy table is not prepared,' it means the fluoroscopy table is not prepared due to the incorrect position of the X-ray tube. To resolve this, confirm the diagnostic table configuration.
What to do if Shimadzu Sonialvision Safire Diagnostic Equipment displays error 'Position Detect Error XXX'?
If the Shimadzu Diagnostic Equipment displays the error message 'Position Detect Error XXX,' it indicates a position detect error in the operating parts. To resolve this, reset by pushing the STOP switch, or re-startup the system.
What to do if Shimadzu Diagnostic Equipment shows error 'I.F. OVER'?
If the Shimadzu Diagnostic Equipment displays the error message 'I.F. OVER,' it means there is an abnormal filament heating current. To resolve this, reduce the set value.
What to do if Shimadzu Sonialvision Safire Diagnostic Equipment shows error 'Starter ERR'?
If the Shimadzu Diagnostic Equipment displays the error message 'Starter ERR,' it indicates the starter is in an abnormal situation or condition. To resolve this, decrease the radiography tube voltage or radiography tube current.
Details the overall system components and layout.
Explains the controls and functions of the diagnostic table console.
Describes the operation and controls of the X-ray high voltage generator.
Outlines the main display and controls of the digital image processor.
Provides a visual guide to the system's operational sequence.
Step-by-step guide for powering on the system and its components.
Instructions for properly closing active studies and powering off the system.
Procedures for calibrating the Flat Panel Detector (FPD) to ensure image quality.
Alternative method to perform FPD calibration using the system's context menu.
Process for entering patient information to initiate a new study.
Method for quickly entering patient data during emergency situations.
How to select the appropriate study type based on imaging needs.
Steps to configure parameters for radiography and fluoroscopy procedures.
Detailed procedure for performing SPOT and SERIAL radiography.
Instructions for performing Digital Subtraction Angiography (DSA) if the option is available.
Guide for operating the digital tomography function, including parameter setup.
Methods for printing acquired images from the system.
Procedures for printing specific images using the context menu.
How to print entire patient files or selected studies from the patient list.
Instructions for transmitting images and studies using DICOM network or CD-R.
Steps to properly close the current patient study.
Procedures for using emergency stop and recovering the system.
Lists common error messages from the X-ray generator and their solutions.
Steps to take for system recovery and verification after a power interruption.