Transport Mode
You can achieve up to 10 hours transport time by
releasing the umbilical cord from the pump unit.
To achieve this press down on the release button
and remove the connector.
The air is sealed in the mattress.
Turn the pump o.
Alarms
The visible and audible alarm function has clear
identification of the fault, which is displayed on the
LED screen to avoid confusion. The alarm can be
muted for 20 minutes but will re alarm to indicate the
fault hasn’t been resolved.
Troubleshooting
Low Pressure
Pressure is not reaching identified levels – check
tube set is connected, CPR has been attached, air
cells not inflating.
High Pressure
Check umbilical cord isn’t trapped.
Power Fail
Check the mains cable is plugged in to the power
supply and turned on. Check that there is no
damage to the mains cable.
System Fault
Check that pump hasn’t been disconnected
during inflation.
Once the fault has been resolved please turn the
pump o and on to reset it.
No Patient Detected:
(only if Egress Alert has been activated)
If the patient gets o the bed the pump will alarm
and indicate on the display that no patient weight is
detected. You can press any button to disable the
alarm. If the mattress is not required i.e. patient is
sat out, it is advisable to disconnect the tube set and
turn the pump o. The mattress will remain inflated.
CPR
Rapid deflation of the mattress may be required for
emergency treatment or system deflation. The CPR
dial is located at the foot end of the mattress.
Rotate the CPR dial to the open position, once done
the entire system will rapidly deflate.
To re-inflate turn the CPR dial to the closed position.
Wait for the mattress system to reach optimal
pressure prior to a return to normal use.
Leaflet Code: LL279
Need Assistance?
Call 01422 233 136 (8:30am - 5pm)
or 0800 037 0234 (out of oce hours)
Quick Glance User Guide
Cable Management
To reduce risk of trip hazards ensure the mains cable
is secured with the cable management along the
side of the mattress.
Always ensure cable is unplugged from mains power
before moving the bed. It is advised not to wrap the
cable tightly but to leave some slack.
Press release button to
remove connector
Thread cable through top end Secure down the length of
the mattress