Other technical specifications
Insulation tests,
type tests
according to IEC 61010-1
Inputs (currents against currents
and voltages)
3.7 kV; 50 Hz; sinusoid;
6.8 kV surge voltage;
1.2/50 µs; R
i
= 500 Ω
Inputs against all analog outputs,
interface, pulse output, and
auxiliary voltage
5.55 kV; 50 Hz; sinusoid;
10.2 kV surge voltage;
1.2/50 µs; R
i
= 500 Ω
Auxiliary voltage against all analog
outputs, interface and pulse output
3.7 kV; 50 Hz; sinusoid;
6.8 kV surge voltage;
1.2/50 µs; R
i
= 500 Ω
Insulation tests,
routine tests
according to IEC 61010-1
Inputs (currents against currents
and voltages)
2.2 kV; 50 Hz; sinusoid
Inputs against all analog outputs,
interface, pulse output, and
auxiliary voltage
3.25 kV; 50 Hz; sinusoid
Auxiliary voltage against all analog
outputs, interface and pulse output
2.2 kV; 50 Hz; sinusoid
Analog outputs, interface and pulse
output mutually
700 V DC
Caution!
Analog output 1 is electrically connected to the interface