Siemens Healthcare Diagnostics
Urgent Field Safety Notice
10813924, Rev. A
November 2012
ADVIA Centaur
®
ADVIA Centaur
®
XP
System Misinterprets Wash 1 Fluid Signals
Our records indicate that you have received the ADVIA Centaur
®
or ADVIA Centaur XP system
(catalog number: 078-A001-xx ADVIA Centaur system, 078-A002R02 ADVIA Centaur
refurbished , 078-A011-03 ADVIA Centaur XP system, 078-A011R03 ADVIA Centaur XP
refurbished).
Reason for Voluntary Correction
Siemens Healthcare Diagnostics is conducting a voluntary field correction for the
ADVIA Centaur and ADVIA Centaur XP system.
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed
occurrences of the system either:
Failing to detect that the Wash 1 bottle and reservoir is empty, or
Incorrectly indicating that the Wash 1 bottle and reservoir is empty when it is actually full.
If the system fails to detect an empty Wash 1 bottle and reservoir it can continue to process
samples and produce erroneous but believable results which are falsely high or low (depending
if the assay architecture is direct or indirect). This issue potentially affects only those assays that
use Wash 1 (Table 1).
Table 1. Assays Potentially Affected
Assay Type Assay
Auto-immune ANA
Bone VitD
Cardiovascular BNP, DDimer, TnIUltra
Torch and Special ID RUBG2, RubM, ToxG, ToxM
Reproductive Endocrinology DHEAS, eE2, SHBG
Infectious Disease
aHAVM, aHAVT/HAVT, aHBcM, aHBcT/HBcT, aHBe, aHBs, aHBs, aHBs2,
aHCV, CHIV, CMVG, Conf, EHIV, HBeAg, HBs, HIV, PCT, SYPH
Immunosuppressant CsA
Liver Fibrosis HA, PIINP, TIMP1
Tumor Markers / Oncology CA153, CA19-9, CA199A, FreePSA
511 Benedict Ave.
Tarrytown, NY 10591 www.siemens.com/diagnostics
Page 1 of 5
Siemens Healthcare Diagnostics All rights reserved. 10813924, Rev. A
GPF-003 -11 V1.0 Effective: September 21, 2012
Related Procedure: GP-003 DX- Field Correction Action

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ADVIA Centaur and is the answer not in the manual?

Siemens ADVIA Centaur Specifications

General IconGeneral
CategoryImmunoassay Analyzer
ThroughputUp to 240 tests per hour
Sample TypesSerum, plasma, urine, CSF
Assay TechnologyChemiluminescence
TypeAutomated
Assay MenuAllergy, Anemia, Cardiac, Fertility, Infectious Disease, Thyroid
Power Requirements200-240 VAC, 50/60 Hz

Related product manuals