Page 114 SITRANS LUT400 (HART) – OPERATING INSTRUCTIONS A5E33329501
mmmmm
General Operation
Palmer Bowlus Flume
Application information
Sized by pipe diameter D
Flume relief is trapezoidal
Designed to install directly into pipelines and manholes
Head is referenced to bottom of the throat, not bottom of the pipe
For rated flows under free flow conditions, the head is measured at a distance of
D/2 upstream from the beginning of the converging section
Parameter Value
2.15.1. Primary Measuring Device (PMD)
Palmer Bowlus Flume
2.15.4. PMD Dimensions
Maximum Flume width h
max
2.15.3.3. Maximum Head
2.15.3.4. Maximum Flow at 20 mA
2.15.3.5. Zero Head Offset
2.15.3.7. Flowrate Units
2.15.3.1. Method of Flow Calculation
Ratiometric
Note: Palmer Bowlus Flume can only be setup using ratiometric calculations
Zero
Head
transducer**
D/2, point of measurement*
Flow
Plan View
Front View
Side View
D = pipe or sewer diameter
* for rated flows under free flow conditions
** The transducer must be above the maximum head by at least the blanking value (see
2.2.4.
Near Range
).
D