PROGRAM
OPTIONS
(CONTINUED)
.NOTE:
The
maximum
values
set
apply
for
all
operating
modes
(Primary,
Piggyback
and
Program
modes).
3.3
To
check
or
change
the
maximum
settings,
with
the
unit
OFF,
hold
the
#5
key
down
while
turning
the
unit
ON.
An
audible
“beep”
will
sound
when
access
has
been
completed.
Release
the
#5
key.
Within
5
seconds,
enter
code
3, 0, 3.
The
unit will
then
display
“SET
MAX”.
3.3.1
To
check
or
change
the
maximum
Rate
setting,
press
the
primary
RATE
key.
The
unit
will
then
display
“MRAT
UUU”
(UUU
=
the
present
maximum
rate
limit).
If
you
want
to
change
the
present
setting,
enter
the
value
using
the
numeric
portion
of
the
keyboard.
NOTE:
The
maximum
Rate
setting
range
is
from
20ml/hr
to
999ml/hr.
To
save
the
new
value
push
the
RUN/STOP
switch.
This
will
store
the
new
value
in
the
non-volatile
memory.
The
unit
will
then
“beep”
and
display
“MRAT
XXX”
(XXX
=
new
value
entered).
3.3.2
To
check
or
change
the
maximum
Volume
setting,
press
the
primary
VOL
LIMIT
key.
The
unit
will
then
display
“MVOL
UUU”
(UUU
=
the
present
maximum
rate
limit).
If
you
want
to
change
the
present
setting,
enter
the
value
using
the
numeric
portion
of
the
keyboard.
NOTE:
The
maximum
Rate
setting
range
is
from
50ml
to
9999ml.
To
save
the
new
value
push
the
RUN/STOP
‘switch.
This
will
store
the
new
value
in
the
non-
volatile
memory.
The
unit
will
then “beep” and
display
“MVOL
XXX"
(XXX
=
new
value
entered).
3.3.3.
After
completing
the
check
and/or
change,
press
OFF.
4.
Low
RATE
limit
5.
4.1
Standard
Operation
The
Rate
can
be
set
as
low
as
0.1mi/hr:
4.2
Optional
The
lower
limit
can
be
changed
to
restrict
the
lower
limit
to
Iml/hr.
This
means
that
the
user
cannot
set
the
Rate
any
lower
than
1ml/hr.
The
user
can
still
use
the
decimal
in
setting
a
Rate
larger
than
1.0ml/hr
(i.e.
1.1,
99.5
etc.).
4.3
To
Check
or
change
this
option,
with
the
unit
OFF,
hold
the
#4
key
down
while
turning
the
unit
ON.
An
audible
“beep”
will
sound
indicating
entry
into
this
mode.
Release
the
#4
key.
Within
5
seconds,
enter
code
9, 5,
1.
The
unit
will
then
display
“LRAT
X_X”
(X.X
=
either
0.1
or
1.0).
To
change
the
present
setting,
press
3,
0, 3.
After
completing
the
selection,
press
OFF.
This
will
store
the
selection
in
the
non-volatile
memory
(battery
backed
RAM).
NOTE:
The
Lower
Rate
Limit
applies
for
all
operating
modes
(Primary,
Piggyback
and
Program
mode).
PIGGYBACK/PROGRAM
Enable
(Disable)
Option
5.1
Standard
Operation
The
Piggyback/Program
keys
can
be
disabled
during
normal
pump
operations.
5.2
To
access
option;
hole
the
#6
key
down
when
the
pump
is
turned
ON.
A
short
audio
“beep”
will
sound
when
operation
is
complete.
The
unit
will
display
“PGBK
ON”
or
“PGBK
OFF”
when
enabled
or
disabled.
5.3
The
option
can
be
toggled
on
or
off
by
entering
the
key
sequence
3, 0,
3
within
5
seconds
of
releasing
the
#6
key.
5.4
“PGBK
ON”
~
indicates
Piggyback/Program
keys
are
enabled.
“PGBK
OFF”
indicates
Piggyback/Program
keys
are
disabled.
5.5
After
completing
the
selection,
press
OFF.
This
will
store
the
selection
in
the
non-volatile
memory
(battery
backed
RAM).
.
Section
1
Page
21