Details the main components of the BS3000S analyzer, including Control System, Optical System, and Printer.
Outlines key specifications like wavelength, light source, absorbency range, and analytical methods.
Specifies the necessary space and environmental conditions for installing the analyzer.
Explains that the analyzer's principle is based on Lambert-Beer Law.
Guides users through the initial screen, test selection, and main interface functions.
Covers setting parameters like temperature display, filter, cuvette, and aspiration.
Details essential operational steps like preheating and pipeline washing.
Explains the AD Auto Zero process for measuring light intensity and obtaining absorbance.
Describes the light scattering principle used for specific proteins testing.
Guides users on selecting scattered light and using cuvettes for specific proteins tests.
Focuses on washing the flowing cuvette daily to maintain cleanliness and performance.
Details the weekly procedure of washing the flowing cuvette with detergent and rinsing.
Covers cleaning dust/stains on the casing and checking peristaltic pump absorption.
Advises on proper handling and use of coagulation and micro detector cuvettes.
Provides measures to eliminate malfunctions and guidance for contacting support.
Lists causes like dirty cuvettes, air bubbles, or faulty pipette.
Lists causes like air bubbles, pump issues, unstable voltage, or invalid reagent.
Identifies fuse issues or poor power connection as common causes.
Attributes screen changes to voltage variations and suggests potentiometer adjustment.
Points to a loose printer head cable as the primary cause.
Cites ambient temperature and insufficient heating voltage as reasons.
Suggests insufficient incubation time or calibration needs.
Explains the need for proofreading and replacing parts for reliable analyzer operation.
Provides step-by-step instructions for safely replacing the instrument's fuse.
Details the procedure for replacing the halogen lamp and LED light sources.
Outlines the steps required to replace the peristalsis pump assembly.
Provides instructions for replacing the printing paper in the analyzer's printer.
Specifies transport regulations, avoiding hazards and environmental conditions.
Details ideal environmental conditions for storing the analyzer to prevent damage.