Sinnowa Medical Science & Technology Co., LtD
BS3000S Multi-test Analyzer
Users Manual
Question and Answer IconNeed help?

Do you have a question about the SINNOWA BS Series and is the answer not in the manual?

Summary

General Description

Configuration and Structure

Details the main components of the BS3000S analyzer, including Control System, Optical System, and Printer.

Main features and technical parameters

Outlines key specifications like wavelength, light source, absorbency range, and analytical methods.

Installation

Installation Requirement

Specifies the necessary space and environmental conditions for installing the analyzer.

Biochemical Test Functions and Operation

Working Principle

Explains that the analyzer's principle is based on Lambert-Beer Law.

Operation

Guides users through the initial screen, test selection, and main interface functions.

Parameters settings

Covers setting parameters like temperature display, filter, cuvette, and aspiration.

Instrument operation key process

Details essential operational steps like preheating and pipeline washing.

AD Auto Zero

Explains the AD Auto Zero process for measuring light intensity and obtaining absorbance.

Specific Proteins Test Function and Operation

Principle

Describes the light scattering principle used for specific proteins testing.

Operation

Guides users on selecting scattered light and using cuvettes for specific proteins tests.

Device Maintenance

Daily Maintenance

Focuses on washing the flowing cuvette daily to maintain cleanliness and performance.

Weekly Maintenance

Details the weekly procedure of washing the flowing cuvette with detergent and rinsing.

Monthly Maintenance

Covers cleaning dust/stains on the casing and checking peristaltic pump absorption.

Other Maintenance Items

Advises on proper handling and use of coagulation and micro detector cuvettes.

Troubleshooting

Malfunction phenomenon and maintenance

Provides measures to eliminate malfunctions and guidance for contacting support.

Malfunction phenomenon: There are mistakes with auto zero

Lists causes like dirty cuvettes, air bubbles, or faulty pipette.

Malfunction phenomenon: Wrong result or bad repetition

Lists causes like air bubbles, pump issues, unstable voltage, or invalid reagent.

The instrument does not work

Identifies fuse issues or poor power connection as common causes.

Screen definition changes

Attributes screen changes to voltage variations and suggests potentiometer adjustment.

Printer keeps walking

Points to a loose printer head cable as the primary cause.

Heating time is long

Cites ambient temperature and insufficient heating voltage as reasons.

Testing time slightly long

Suggests insufficient incubation time or calibration needs.

The corrections and replacements to common parts of analyzer

Explains the need for proofreading and replacing parts for reliable analyzer operation.

The replacements of fuse

Provides step-by-step instructions for safely replacing the instrument's fuse.

The replacement to light source

Details the procedure for replacing the halogen lamp and LED light sources.

The replacement to peristalsis pump

Outlines the steps required to replace the peristalsis pump assembly.

Replacement to printing paper

Provides instructions for replacing the printing paper in the analyzer's printer.

TRANSPORTS AND STORAGE

Transport

Specifies transport regulations, avoiding hazards and environmental conditions.

Storage

Details ideal environmental conditions for storing the analyzer to prevent damage.