Questions and Answers

  • S
    Sonya HarrisonAug 9, 2025
    What to do if SINO SN-50 Water Pump doesn't respond when pressing the power button?
    • M
      Michael CookAug 9, 2025
      If your SINO Water Pump does not respond when you press the power button, it could be due to a low battery and no AC power supply connected. Connect the AC power supply and keep charging the battery. If this does not solve the issue, the power system may have a failure, and you should send it back to the factory.
  • A
    Adam CannonAug 19, 2025
    Why does the “Occlusion” alarm frequently occur in the infusion process of SINO Water Pump?
    • V
      Vanessa WilliamsAug 20, 2025
      Frequent “Occlusion” alarms during the infusion process of your SINO Water Pump can occur if the infusion apparatus pipeline is tied or the roller clamp isn't open. Recheck the pipeline of the infusion apparatus. Alternatively, the pressure level might be set too low, so increase the pressure level. If the issue persists, there may be a failure in the pressure testing system.
  • J
    jenniferlambAug 28, 2025
    How to fix “Drip Error" message that appears frequently when the drip sensor alarm function is enabled on SINO SN-50?
    • R
      robertperezAug 29, 2025
      If the “Drip Error" message appears frequently when the drip sensor alarm function is enabled on your SINO Water Pump, it may be due to incorrect installation of the drip sensor. Reinstall the drip sensor correctly. It could also be that the infusion apparatus is inappropriate or not calibrated properly; select a calibrated infusion apparatus or recalibrate the infusion apparatus.

Summary

1. SAFETY INFORMATION

1.1 Stipulations and Overview

Manual stipulations and general safety overview for the infusion pump.

2. PRODUCT INTRODUCTION

2.2 Technical Specs and Connectivity

Technical characteristics, parameters, RS232 interface, and multi-channel injection.

3. OPERATING METHODS

3.1 Installation and Power Operations

Procedures for pump installation, power on/off, and system self-test.

3.2 Infusion Setup and Modes

Connecting components, selecting modes, and setting infusion parameters.

4. ALARM PRESENTATION

5. MALFUNCTION AND TROUBLESHOOTING

Troubleshooting Guide

Guide to diagnose and resolve common malfunctions and error messages.

6. MAINTENANCE

SINO SN-50 Specifications

General IconGeneral
ModelSN-50
CategoryWater Pump
Power50W
Max Head2.5M
Voltage220-240V
Frequency50Hz
Cable Length10M
TypeSubmersible