Operating Instruction SIROSONIC/L/TL, PerioSonic 5Operation
62 78 415 D3275.201.04.09.02 63
5.7.7 Using the SIROIMPLANT tip
CAUTION! SIROIMPLANT IP 1 may only be used on SIROSON L,
SIROSONIC/L/TL and PerioSonic.
This tip is used for biofilm management and for cleaning implants as
well as metallic and ceramic restorations.
Prior to every use of the tip:
● Check to make sure that no metal is visible underneath the coating.
If this is the case, replace the cleaning insert.
● Tighten the holder of the cleaning insert with the knurled bushing
and check the cleaning insert to make sure that it is firmly attached.
We recommend placing a cofferdam to prevent swallowing or inhalation
of a detached or broken-off fragment. If it is not possible to place a
cofferdam, the patient should breathe through his or her nose.
Work with a maximum rinsing medium flow rate in order to ensure
sufficient cooling of the plastic sheathed cleaning pin.
Insert IP 1 can be used and sterilized repeatedly.