8Maintenance Operating Instruction SIROSONIC/L/TL, PerioSonic
76 62 78 415 D3275.201.04.09.02
8.3 Replacing the spray water cartridge
Required accessories
● Spray water cartridge
T1 spray
● Sirona T1 Spray
If the spray water cartridge leaks or is clogged.
SIROSONIC/L
... for SIROSONIC/L
Replacing spray in SIROSONIC LScaler
1. Detach the ultrasonic handpiece from the supply hose.
2. Detach the control ring from the hose coupling.
3. Remove the spray water cartridge (A).
4. Spray oil lightly onto the new spray water cartridge.
5. Insert the new spray water cartridge up to the stop.
6. Attach the control ring with the smooth ring side first onto the supply
hose.