7Conditioning Operating Instruction SIROSONIC/L/TL, PerioSonic
72 62 78 415 D3275.201.04.09.02
7.4 Sterilizing
Steps _ Scaler I I
✔ The instruments are cleaned and disinfected.
✔ If necessary, the ultrasonic handpiece and accessories can be
sterilized in packaging suitable for sterilization and storage, e.g.
paper/plastic composite packaging or container.
➢ Sterilize the ultrasonic handpiece and accessories in the steam
sterilizer with saturated water vapor.
Unit pressure in bar and psi ww
Overpressure: 2.04 bar (29.59 psi)
Temperature: 134° C (274° F)
Holding time: 3 min
6278415.book Seite 72 Montag, 17. Oktober 2016 11:28 11