Operating Instruction SIROSONIC/L/TL, PerioSonic 7Conditioning
62 78 415 D3275.201.04.09.02 73
DAC PROFESSIONAL _ Scale r
Steam sterilizers that comply with the requirements of either EN 13060,
class B (for example, DAC PREMIUM / DAC PROFESSIONAL) or
EN 13060, class S and are also suitable for the sterilization of ultrasonic
handpieces and ultrasonic tips are approved.
NOTICE! Do not exceed 140°C (284°F), even during the drying phase.
After sterilizin g - heading
After sterilizing
Next steps _ Instruments (plural)
1. Remove the instruments from the steam sterilizer immediately.
CAUTION! The instruments are hot. Risk of burns!
NOTICE! Do not attempt to accelerate the cooling process by
immersing the instruments in cold water. This can damage your
instruments!
2. Store all instruments so that they are protected from contamination.
3. Sterilize the instruments again once the storage period has elapsed.