OPERATION
12. If the operator elects to abort a test in process this can be accomplished by pressing
the RESET switch at anytime. The instrument will stop the test process immediately and
the display will show:
A b o r t MX XXX . X s
X . XX K V A C XX . XX mA
W - A B R T MX XXX . X s
X . XX K V A C XX . XX mA
Please press the TEST switch to initiate another test.
Insulation Resistance Test Mode model 2945
13. To initiate a test press the TEST switch on the front panel. The red High Voltage
Arrow indicator will flash and the display will immediately show:
I - T e s t MX XXX . X s
X . XX K V D C XXX . X M
14. If the DUT passed the test, the instrument will output a short audible beep tone to
indicate the DUT has passed the test. The display will show:
I -P a s s MX XXX . X s
X . XX K V D C XXX . X M
15. If the operator elects to abort a test in process this can be accomplished by pressing
the RESET switch at anytime. The instrument will stop the test process immediately and
the display will show:
I - A B R T MX XXX . X s
X . XX K V D C XXX . X M