EasyManua.ls Logo

Slaughter 2925 - OPTIONS

Slaughter 2925
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION
38
12. If the operator elects to abort a test in process this can be accomplished by pressing
the RESET switch at anytime. The instrument will stop the test process immediately and
the display will show:
models 2925 and 2935
A b o r t MX XXX . X s
X . XX K V A C XX . XX mA
model 2945
W - A B R T MX XXX . X s
X . XX K V A C XX . XX mA
Please press the TEST switch to initiate another test.
Insulation Resistance Test Mode model 2945
13. To initiate a test press the TEST switch on the front panel. The red High Voltage
Arrow indicator will flash and the display will immediately show:
I - T e s t MX XXX . X s
X . XX K V D C XXX . X M
14. If the DUT passed the test, the instrument will output a short audible beep tone to
indicate the DUT has passed the test. The display will show:
I -P a s s MX XXX . X s
X . XX K V D C XXX . X M
15. If the operator elects to abort a test in process this can be accomplished by pressing
the RESET switch at anytime. The instrument will stop the test process immediately and
the display will show:
I - A B R T MX XXX . X s
X . XX K V D C XXX . X M

Related product manuals