OPERATION
Please press the SET key until the display shows:
Models 2925, and 2935 Model 2945
Use the Up/Down Arrow keys to enter the leakage current high limit setting, then press
the EXIT key to exit to the test mode or toggle to another setting using the SET key. The
unit of measure is in milliamperes with 0.10 AC and 0.02 DC milliamperes as the
minimum setting.
6. To set the Low Leakage Current Limit on models 2925, 2935, and 2945
Please press the SET key until the display shows:
Models 2925, 2935 Model 2945
Use the Up/Down Arrow keys to enter the leakage current low limit setting, then press the
EXIT key to exit to the test mode or toggle to another setting using the SET key. The unit
of measure is in milliamperes with 0.00 milliamperes AC and DC as the minimum
setting.
7. To set the Ramp Time on models 2925, 2935, and 2945
Please press the SET key until the display shows:
Models 2925, 2935 Model 2945
Use the Up/Down Arrow keys to enter the ramp time setting, then press the EXIT key to
exit to the test mode or toggle to another setting using the SET key. If the ramp time is
set to “0”, the instrument will raise the voltage up to the test voltage in a fixed ramp time
of 0.1 seconds AC and DC and then begin the dwell cycle. The minimum ramp time
setting is 0.2 seconds with 0.1 second resolution.
8. To set the Dwell Time
Please press the SET key until the display shows:
Models 2925, and 2935 Model 2945
W - D w e l l = XXX . X s
Use the Up/Down Arrow keys to enter the dwell time setting, then press the EXIT key to
exit to the test mode or toggle to another setting using the SET key. If the dwell time is