Manual_MPS_2021-10-022.00_EN.docx 7/28
5. SPECIFICATIONS
MPS P MPS P I MPS W MPS C MPS CX
115 V 230 V 115 V 230 V 115 V 230 V 115 V 230 V 115 V 230 V
Nominal supply
voltage
115 V AC
+/-10 %
230 V AC
+/-10 %
+/-10 %
230 V AC
+/- 10 %
115 V AC
+/-10 %
230 V AC
+/-10 %
+/-10 %
+/-10 %
+/-10 %
+/-10 %
Nominal frequency
60 Hz 50 / 60 Hz 60 Hz 50 / 60 Hz 60 Hz 50 / 60 Hz 60 Hz 50 / 60 Hz 60 Hz 50 / 60 Hz
Power draw
400 VA 650 VA 320 VA 150 VA 250 VA
Protective Class (1)
I
Power fuses
2xT4 A 2xT2,5 A 2xT10 A 2xT4 A 2xT4 A 2xT2,5 A 2xT10 A 2xT4 A 2xT10 A 2xT4 A
Pollution degree
(1)
2
Overvoltage
installation
category
II
Maximum heat
emission
400 J / s 650 J / s 320 J / s 150 J / s 450 J / s
Operating
temperature range
+10 to +35 °C
Temperature
range during
storage
+5 to +55 °C
Relative humidity max. rel. 80 % non-condensing
Humidity during
storage
max. rel. 80 % non-condensing
Dimensions
(L*W*H) [mm]
450 x 600 x 290 450 x 600 x 290 300 x 600 x 290 400 x 600 x 290 600 x 600 x 290
Weight [kg] 21 23 15 25 36
Paraffin tank 5.5 l 3.8 l
Cooling spot Peltier Peltier
Forceps holders 2 x 4 2 x 3
Light LED LED
Temperature
range cool plate
up to -15 °C
programmable
up to -15 °C
programmable
*
MPS P MPS P I MPS W MPS C MPS CX
Cooling gas R 134 a R 134 a