Manual_MPS_2021-10-022.00_EN.docx 9/28
6. UNPACKING AND INSTALLATION
6.1 UNPACKING THE DEVICE
Remove the upper cover and the upper supporting foams.
Lift the device out of the wooden transportation case. The devices may only be lifted
holding them at the sides of the base plate of the housing.
For repacking use the original cases. Keep the packing material.
Place the device onto the selected bench.
It must be avoided to tilt or turn the cool plate MPS C and MPS CX.
These positions will inevitably cause damage to the compressor.
6.2 INSTALLATION
The site for installation should meet the following requirements:
The unit should be positioned onto a plane, vibration-free surface. Please
ensure that there is no air condition vent nearby.
A free ventilation of fresh air from underneath the device should be
guaranteed. The back of the device must be at least 15 cm away from the
wall.
The vicinity of the work area must be free of oil and chemical vapors.
6.3 ELECTRICAL POWER CONNECTION
Make sure that electric power is constant +/-5 - 10 %:
• This should be examined during installation of the unit by a
competent person.
• Use a dedicated fuse for the unit.
• Before turning on the device, check if the voltage of the mains
supply is identical with the name plate of the unit.
Connect each unit to the mains supply and switch on with mains switch at the rear of
each unit.