HOTLINE
®
Blood and Fluid Warmer Technical Service Pack
55
SECTION 5
Maintenance Solutions
■ Unplug the HOTLINE®.
■ (Old case) Remove the Fill Port plug and hold the Unit over a
sink to empty the water.
(New case) Deploy the drain tube to empty the Unit.
■ Rinse reservoir with distilled water twice.
■ Refill the reservoir with 1.4 litres of disinfectant flushing
solution made to one of the formulae shown in the blue
panel above. Do not fill the HOTLINE® reservoir with a
Disposable Set in place, as this may result in an airlock in the
HOTLINE® warmer.
■ Connect a Disposable Set to the HOTLINE®.
■ Plug the Unit in and switch on.
■ Allow the disinfectant solution to circulate for 30 minutes.
Drain Tube
(in deployed posn)
Drain Tube End Cap
Fill Port
Pole Clamp
Interlock Block
disposable connector
Stowage Clip
(for drain tube when
not deployed)
GM-HL90-8508-en
Figure 5 - 4: Rear view of new-style HOTLINE case, showing drain tube