HOTLINE
®
Blood and Fluid Warmer Technical Service Pack
46
SECTION 5
Maintenance
Maintenance
Before each use
Carefully inspect the HOTLINE
®
for signs of damage, cracked or
split case, insecure poleclamp, damaged mains cable, etc.
If the Disposable Set does not install easily, lubricate the O-
Ring Seals.
Lubricating the O-Ring Seals
Using a cotton swab, apply a small amount of silicone grease
(Smiths Medical Part Number EZL 80-04-002) to the O-Rings in the
Disposable Set connector block.
If this does not solve the problem, reject the unit.
After each use
Wipe all external surfaces of the HOTLINE
®
with a soft cloth,
using an aqueous solution of mild detergent. If necessary, the
external surfaces may be disinfected using a solution of 10% bleach
in distilled water.
Do not autoclave.
Do not use alcohol or solvents.
Do not use abrasive cleaning agents.
Do not use cold sterilants.
Do not immerse any part of the HOTLINE
®
in liquids.