HOTLINE
®
Blood and Fluid Warmer Technical Service Pack
69
SECTION 5
HL-90 Service Test Specification
Other Territories
It is common practice both in the U.S.A. and in the U.K . to set
a nominal temperature of 42°C.
In other territories, a different set point (40°C) may be adopted
as the “normal” setting.
For reference, the details for the alternative Nominal
Temperature adjustment set points are shown here.
Nominal Temperature Water Temperature
(paragraph 6)
Over Temperature
Set Point (paragraph 7)
40 39.9 41.1
42 41.9 43.1