HOTLINE
®
Blood and Fluid Warmer Technical Service Pack
48
SECTION 5
Every 30 days
Every 30 days
Perform a Visual Inspection looking for damage to the case,
poleclamp, mains cable, etc.
Lubricate O-Ring Seals with a cotton swab, applying a small
amount of silicone grease to the O-Rings.
Either, if using DISTILLED WATER as the
recirculating solution:
Drain and replace the recirculating solution as described on
page 54.
or, if using ISOPROPYL ALCOHOL as the
recirculating solution:
Drain and replace the recirculating solution as described on
page 56.
Finally
Fill in, sign, and date the maintenance log record!