HOTLINE
®
Blood and Fluid Warmer Technical Service Pack
51
SECTION 5
Every 12 Months
Observe that:
the GREEN LED extinguishes.
the RED recirculating Solution Level LED lights up.
the AUDIBLE ALARM commences beeping.
the RECIRCULATING SOLUTION ceases circulating.
If any of the above responses are missing, the unit must be
removed from service and repaired. The unit should return to
normal operation upon releasing the float switch.
Disposable Set Interlock Alarm
While the unit is working normally, gently ease the Disposable
Set away from the interlock block. Be aware that a small amount of
recirculating solution may escape during this test. Before the
Disposable Set is fully disconnected, the alarm should operate.
Observe that:
the GREEN LED extinguishes.
the RED Interlock Alarm LED lights up.
the AUDIBLE ALARM commences beeping.
the recirculating SOLUTION ceases circulating.
If any of the above responses are missing, the unit must be
removed from service and repaired. The unit should return to
normal operation upon correctly reattaching the Disposable Set.