HOTLINE
®
Blood and Fluid Warmer Technical Service Pack
53
SECTION 5
Every 12 Months
Apply a smear of silicone grease (Smiths Medical Part
Number EZL 80-04-002) to two new O-Rings, and locate
them in the sockets.
Re-attach the disposable interface block.
Re-fit the socket head screws. Again, take care not to
damage or bend the microswitch operating lever.
A kit of parts is available to make this task simpler. Please ask for
part number reference EZL 80-04-001.
Either, if using DISTILLED WATER as the
recirculating solution:
Drain and replace the recirculating solution as described on
page 54.
or, if using ISOPROPYL ALCOHOL as the
recirculating solution:
Drain and replace the recirculating solution as described on
page 56.
or, if using HYDROGEN PEROXIDE as the
recirculating solution:
Drain and replace the recirculating solution as described on
page 57.
Finally
Fill in, sign, and date the maintenance log record!