DSTA-40
TEMP
CHECK
Thermometer
Operator’s
Manual
7
Repeat
step
4,
step
5,
and
step
6
until
the
Over
Temperature
Alarm
activates
at
43.9°C.
8
Once
the
Over
Temperature
activation point
is
set,
adjust
the
H-1200
Fluid
Warmer
DISPLAY
to
equal
the
TEMP
CHECK
display.
9
Press
the
H-1200
Fluid
Warmer
display
label
back
into
position.
10
Verification
is
complete.
Step
11:
Replace
cover
1
Turn
OFF
the
H-1200
Fluid
Warmer.
2
Unplug
from
the
power
outlet.
3
Remove
the
TEMP
CHECK.
4
Attach
the
rear
panel
ground
wire.
5
Align
the
screw
holes
and
red
air
supply
port
with
the
holes
on
the
back
cover.
6
Using
a
#3
Phillips
head
screwdriver,
replace
all
18
screws
to
secure
the
back
cover.
7
Secure
the
H-1200
Fluid
Warmer
to
the
IV
pole.
8
Insert
the
black
pressure
tube
firmly
into
the
air
supply
port.
Limited
Warranty
Smiths
Medical
ASD,
Inc.
(the
“Manufacturer”)
warrants
to
the
Original
Purchaser
that
the
TEMP
CHECK
Thermometer
(TEMP
CHECK),
not
including
accessories,
shall
be
free
from
defects
in
materials
and
workmanship
under
normal
use,
if
used
in
accordance
with
this
Operator’s
Manual,
for
a
period
of
one
year
from
the
actual
date
of
sale
to
the
Original
Purchaser.
THERE
ARE
NO
OTHER
WARRANTIES.
This
warranty
does
not
cover
normal
wear
and
tear
and
maintenance
items,
and
specifically
excludes
hoses,
warming
blankets,
roll
stands
or
any
other
accessory
items
or
equipment
used
with
the
TEMP
CHECK.
Subject
to
the
conditions
of
and
upon
compliance
with
this
Limited
Warranty,
the
Manufacturer
will
repair
or
replace
at
its
option
without
charge
(except
for
a
minimal
charge
for
postage
and
handling)
any
TEMP
CHECK
(not
including
accessories)
which
is
defective
if
a
claim
is
made
during
such
one-year
period.
The
following
conditions,
procedures,
and
limitations
apply
to
the
Manufacturer's
obligation
under
this
warranty:
A.
Parties
Covered
by
this
Warranty:
This
warranty
extends
only
to
the
Original
Purchaser
of
the
TEMP
CHECK.
This
warranty
does
not
extend
to
subsequent
purchasers.
The
Original
Purchaser
may
be