Check installation of the new collection container. If the
vacuum is low there is a possibility that the o-rings on the
collection container did not seal into the receivers. With the
vacuum running, remove the collection container, check
the o-rings and re-install per the collection container
installation procedure.
Yes:
If you have a solids collection cup (installed by
others) - check & clean or replace element/screen
if necessary.
No:
Collection Container
Maintenance
Line rinsing & vacuum line cleaners - cleaners should be:
Non-foaming, de-odorizing, sanitizing & pH between 6 and 10.
Using line cleaners outside of this pH range will void the warranty.
For a complete list of recommended cleaners, visit our website
at www.solmetex.com.
With Dental Vacuum Systems: Follow manufacturers instructions.
Plan to rinse no more than 5 chairs every 10 minutes. Limit
the total maximum flow to the Hg5
®
Mini during rinsing to 1
liter per minute. Rinsing too much or too rapidly can overfill
the Hg5
®
Mini and could aect the unit’s eciency.
Service
The Hg5
®
Mini is designed to provide years of trouble-free service,
with minimal attention. In the unlikely event of system related
problems, please review the Q&A or troubleshooting section.
Q&A: Little or no vacuum at the hand piece.
1. Check sediment level of collection container, if full change the
collection container.
2. If the vacuum gauge reads normal but there is little or no
vacuum to the hand piece, there is probably a clog or a leak
between the hand piece and the Hg5
®
Mini.
3. If the vacuum gauge reads lower than normal these are the
possible causes:
Did you just replace the collection container?
Your collection container must be changed once every
twelve (12) months or when the sediment level reaches
the full line, whichever occurs first.
Check the sediment level of the collection container weekly.