Adjust the blade-to-orifice distance by turning the wheel to move the blade further away from the orifice. If this
increases the acoustic meter reading, continue moving the wheel in the same direction until the reading peaks
on the meter. If the reading decreases, reverse the procedure until the reading is peaked.
Return to the back pressure valve and adjust it slightly for the maximum meter reading.
It is not necessary to reach and maintain the exact maximum meter reading, for the instrument is extremely
sensitive and relatively minor changes are greatly amplified with the result that they appear more significant than
they actually are. Fluctuations in the meter reading during operation are normal and do not require re-tuning of
the Sonolator.
8. With the correct and desired pressure and a proper tune, you are ready to begin running product. Connect your
inlet and outlet piping.
9. Open all valves both upstream and downstream of the Sonolator System.
10. Start unit.
11. If a feed pump is connected to force feed the pump provided with the Sonolator System, make sure to start
the Sonolator System pump first, then slowly pump material with the feed pump carefully observing the inlet
pressure created between the two pumps; this pressure should never exceed 50 to 60 psi.