EasyManuals Logo

Sonoscape A6 User Manual

Sonoscape A6
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
D-2
Table 1-2. Transducer Model: 6-2MHz C351 Curved Array
Operating Mode: THI-mode
TIS
Non-scan
Index
label
MI
Scan
A
aprt
1c A
aprt
>1cm
TIB
Non-
scan
TIC
Maximum index value
0.54 0.11
# # # #
p
r, α
(MPa)
1.02
P (mW)
6.3
# # #
min of (mW)
#
z
s
(cm)
#
z
bp
(cm)
#
z
b
(cm) 4.7 #
z at max I
pi.α
(cm) 4.7 #
d
eq
(z
b
) (cm) #
f
awf
(MHz)
3.6 3.6
# # # #
Dim of A
aprt
X (cm) 1.9 # # # #
Associated
acoustic
parameters
Y (cm)
1.3 # # # #
t
d
(µsec)
0.476
prr (Hz)
4219
p
r
at max I
pi
(MPa) 1.82
d
eq
at max I
pi
(cm) #
I
pi.α
at max MI (W/cm
2
) 25.6
FL
x
(cm) 5 # # #
Other
information
Focal
Length
FL
y
5 # #
#
Mode
THI THI
# # #
#
Focus
(cm)
5 5
# # #
#
PRF
(kHz)
4 4
# # # #
Operating
control
conditions
Power
(%)
100 100
# # #
#

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sonoscape A6 and is the answer not in the manual?

Sonoscape A6 Specifications

General IconGeneral
BrandSonoscape
ModelA6
CategoryMedical Equipment
LanguageEnglish

Related product manuals