EasyManuals Logo

Sonoscape A6 User Manual

Sonoscape A6
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
D-11
Table 4-2. Transducer Model: 11-5MHz L745 Linear Array
Operating Mode: THI-mode
TIS
Non-scan
Index label MI
Scan
A
aprt
1c A
aprt
>1cm
TIB
Non-
scan
TIC
Maximum index value
0.50 0.19
# # # #
p
r, α
(MPa)
1.02
P (mW)
9.1
# # #
min of (mW)
#
z
s
(cm)
#
z
bp
(cm)
#
z
b
(cm) 1.5 #
z at max I
pi.α
(cm) 1.5 #
d
eq
(z
b
) (cm) #
f
awf
(MHz)
4.3 4.3
# # # #
Dim of A
aprt
X (cm) 1.2 # # # #
Associated
acoustic
parameters
Y (cm)
0.6 # # # #
t
d
(µsec)
0.464
prr (Hz)
7692
p
r
at max I
pi
(MPa) 3.09
d
eq
at max I
pi
(cm) #
I
pi.α
at max MI (W/cm
2
) 31.4
FL
x
(cm) 2 # # #
Other
information
Focal
Length
FL
y
(cm)
2 # #
#
Mode
THI THI
# # #
#
Focus
(cm)
2 2
# # #
#
PRF
(kHz)
8 8
# # # #
Operating
control
conditions
Power
(%)
100 100
# # #
#