EasyManua.ls Logo

Sony D-E700 - Page 10

Sony D-E700
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
– 10 –
Focus/Tracking Gain Adjustment
A servo analayzer is necessary in order to perform this adjustment
exactly.
However, this gain has a margin, so even if it is slightly off,
there is no problem. Therefore, do not perform this adjustment.
Focus/tracking gain determines the pick-up follow-up relative to
mechanical noise and mechanical shock when the 2-axis device
operates. However, as these reciprocate, the adjustment is at the
point where both are satisfied.
When gain is rased, the noise when 2-axis device operates in-
crease.
When gain is lowered, it is more susceptible to mechanical shock
and skipping occurs more easily.
This adjustment has to be performed upon replacing any of the
following parts.
Optical pick-up
RV503 (Focus gain VR)
RV502 (Tracking gain VR)
Normally, be sure not to move RV503 (focus gain VR) and
RV502 (tracking gain VR).
Adjustment method:
– Focus Gain Adjustment –
This adjustment is not performed.
If focus gain VR RV503 is turned, set to mechanical center.
Tracking Gain Adjustment –
(perform at normal operation)
1. Place the optical pick-up level, horizontally. (If the optical
pick-up is not level, the 2-axis device will be weighted and
adjustment cannot be done.)
2. Connect the oscilloscope to TP (TE) and TP520 (VC) on MAIN
board.
3. Set the disc (YEDS-18) and press the ^ key.
4. Turn RV502 slightly clockwise (tracking gain drops) and ob-
tain a waveform with a fundamental wave (waveform has large
waves) as in Figure 1.
5. Turn RV502 slowly counterclockwise (tracking gain rises) until
the fundamental wave disappears (no large waves) as in Fig-
ure 2.
6. Set RV502 to the position about 30˚ counterclockwise form
the position obtained in step 5. If RV502 contact point is more
than 90˚ counterclockwise from mechanical center, tracking
gain is too high. In this case, readjust from step 4.
7. Press + or = key and observe the 100 track jump wave-
form. Check that no traverse waveform appears for both +
or = directions. (See Figures 3 and 4.) It is acceptable if
the traverse waveform appears only now and then, but if it
appears constantly, raise tracking gain slightly and check step
7 again.
8. Check that there is not abnormal amount of operation noise
(white noise) from the 2-axis device. If there is, tracking gain
is too high, readjust starting with step 4.
The waveforms are those measured with the oscilloscope set as
shown below.
VOLT/DIV: 500 mV
TIME/DIV: 5 ms
Waveform when tracking gain is lowered.
Fundamental wave appears (large waves).
Fig. 1
Waveform when fundamental wave disappears (no large waves).
Fig. 2
Waveform with no traverse waveform during 100 track jump.
(Brake application is smooth because of adjustment.)
Fig. 3
Waveform with traverse waveform during 100 track jump.
(Brake application is poor because of adjustment.)
Fig. 4
100 track
j
ump waveform
100 track jump
waveform
traverse waveform
oscilloscop
e
(DC range)
MAIN board
TP (TE)
TP520 (VC)
(See pa
g
e 11)
+
10 k
0.0022
µ
F
–– 0 V
–– 0 V
–– 0 V
–– 0 V

Other manuals for Sony D-E700

Related product manuals