5-1-3. RS-232C model
(only LT20-101C/201C)
Proceeds to the next setting mode from 5-1-1. Basic settings
step 4.
1. Setting the output data format
: Normal output
1st byte : Channel name (A or B)
2nd byte : Sign (space: +, “–”)
3rd to 9th bytes : Numerical data (ex.123.456)
: Outputting with measurement mode information
1st byte : Channel name (A or B)
2nd byte : Current mode
( N : Current value,
P : Peak-to-peak value,
I : Minimum value,
A : Maximum value)
3rd byte : Unit (M: mm, I: inch)
4th byte : Sign (+ or “–”)
5th to 11th bytes : Numerical data (ex.000.000)
: Outputting according to the mode 1 format
(statistical calculations) of the digital printer
P40 (P40 is sold separately).
Whether, with the 2 channel model, to output B channel data
following a space or to divide it with the delimiter is selected
by step 9. (except for P40 mode)
Note
When set to
mode even the
2 channel model only outputs the A
channel.
factory-set
5-1-2. BCD model
(only LT20-101B/201B)
Proceeds to the next setting mode from 5-1-1. Basic settings step 6.
1. BCD logic
Setting the BCD output logic.
+ is true logic.
“–” is false logic.
Exception : Logic for the DRQ, READY, and alarm
terminals cannot be changed.
(See P63 7. BCD output)
2. BCD output format
Setting the BCD output format.
:BCD is output
according to the DRQ
signal input, and the
resulting status is
held even if the DRQ
signal goes off.
: BCD is output according
to DRQ signal input,
and assumes high-
impedance status
when there is no DRQ
signal input.
Initial settings are now complete for the BCD model.
Press
MODE
to return to the measuring state.
factory-set
factory-set