LOW-TEMPERATURE BATH OPERATORS MANUAL
FTS
S
YSTEMS
B
IO
C
OOL
80
Controlled Rate Freezing Bath

Questions and Answers

  • A
    Alexander WalkerAug 11, 2025
    How to troubleshoot SP Scientific FTS Systems BioCool 80 that will not start?
    • D
      Dr. Douglas KellyAug 11, 2025
      If your SP Scientific Laboratory Equipment unit won't start, check the power source to ensure it's properly connected and that the voltage matches the unit's specifications. After a power shutdown, wait five minutes before restarting the unit.
  • D
    Daniel RobinsonAug 17, 2025
    What to do if my SP Scientific FTS Systems BioCool 80 unit will not cool?
    • K
      Kenneth BrennanAug 17, 2025
      If your SP Scientific Laboratory Equipment unit isn't cooling, make sure it has enough airflow. The unit pulls air from the left to cool its refrigeration system, which then exits from the rear. Also, check that the Chill light is on. To ensure proper airflow and prevent damage to the refrigeration system, position the unit with at least six inches of clearance on all sides.
  • J
    Jacob HowardAug 23, 2025
    What to do if compressors will not start in SP Scientific FTS Systems BioCool 80 Laboratory Equipment?
    • R
      Robert RichardsonAug 23, 2025
      If the compressors in your SP Scientific Laboratory Equipment won't start, wait five minutes after the last power shutdown before restarting. The BioCool™ 80 refrigeration system compressor has a sensor that prevents the low-stage compressor from starting for several minutes after a power loss, whether intentional or unintentional.
  • C
    Chad JensenAug 29, 2025
    How to fix a magnetic stirrer that does not rotate in SP Scientific FTS Systems BioCool 80 Laboratory Equipment?
    • A
      Amy MendezAug 29, 2025
      If the magnetic stirrer in your SP Scientific Laboratory Equipment isn't rotating, here's what you can do: * Switch off the unit. * Unplug the unit and remove the cover. The driving magnet for the stirrer is located under the bath and is driven by a toothed drive belt attached to a motor. Check the drive belt for proper tension, as it may have tightened or loosened during shipment. If the drive belt has slight tension, loosen the drive motor mounting nuts and slide the motor until the belt shows slight tension (i.e., 3/8- to 1/2-inch of play). Retighten the mounting nuts once you have corrected the drive belt tension, and replace the cover. The drive belt is toothed and does not require full tension for operation.
  • P
    Patricia LewisSep 4, 2025
    How to fix a noisy magnetic stirrer in SP Scientific FTS Systems BioCool 80?
    • D
      Daniel JohnsonSep 4, 2025
      If the magnetic stirrer in your SP Scientific Laboratory Equipment is making noise, ensure that the stirrer is centered in the bath and that the holding screws on the magnetic stirrer are snug.