22 23
• Check with sling label for washing condition.
• As for minor cleaning between patient uses, we recommend wipes with
water-soluble soap or isopropyl alcohol.
• For slings with rigid supports, remove the plastic inserts before washing. (Look
for an opening near the plastic inserts for removal)
• Machine wash warm or cold.
• Do not wash with other colors.
• Air dry or very low dryer heat.
Deterioration of sling
• Contact Span-America for “Guideline for Identifying Deteriorated Slings” if needed.
Disinfecting of the lift and sling
• Inspect slings prior to each use for contamination from previous use.
• Use isopropyl alcohol for any disinfection purpose.
Cleaning and Disinfecting
Cleaning of the lift
Cleaning of the sling
• Solvents or strong liquids or abrasive cleaning products will damage the
surfaces of the lift.
• Use pH-neutral detergents only.
• Remove the battery from the controller
• Clean lift surfaces, control box and hand control with a damp cloth using
pH-neutral detergent.
• Remove debris and hair from the casters and make sure the casters can
spin and swivel freely.
• DO NOT USE BLEACH.
• DO NOT clean or wash disposable slings.
• REPLACE a sling when it show signs of deterioration.
• Wash a reusable sling for serious contamination
• Replace a disposable sling for any contamination
• Reusing a contaminated sling can cause serious bacterial infection
Emergency Lowering Mechanism
Contact your dealer immediately if standard troubleshooting techniques do not
correct the failure. Do not attempt to lift until all failure and safety issues have
been resolved.
In case of lift failure, please follow the procedures below to safely lower the user.
The Emergency Lowering Device is located at the top of the actuator shaft. It is
intended for use if the actuator or electronics fail to operate while the patient is
suspended in midair.
The device consists of a plastic collar ring that should be rst pulled down and turned
clock-wise continually until the patient has been lowered.
NOTE:
THE EMERGENCY LOWERING DEVICE IS INTENDED FOR USE DURING LIFT
FAILURE. THIS DEVICE WILL ALLOW LOWERING OF PATIENTS ONLY. PLEASE
CONTACT YOUR DEALER IMMEDIATELY IN CASE OF FAILURE.